View PDF Format

Legacy

Learn More
Learn More

Stirling SU780XLE

Stirling VAULT100

Stirling VAULT100 PRO

Storage Capacity – 780L

Temperature Uniformity @ -80°C – 9.5°C (9 probes, 3″ from wall)

Temperature Uniformity @ -70°C – 7.0°C (9 probes, 3″ from wall)

Recovery from Door Opening to Steady-State* – 35 min @ -80°C

Warm-Up Time from -80°C to -60°C – 2.5 hrs.

Steady-State Energy Consumption* – 6.67kWh/day at -75°C

Steady-State Heat Dissipation – 981 BTU/h @ -80°C

7″ Touchscreen

Onboard Predictive Analytics – No

BACnet I/P – Yes

External Communications – Dry Contacts, 420mA

Engine – M600

Storage Capacity – 795L

Temperature Uniformity @ -80°C – 6.0°C (9 probes, 3″ from wall)

Temperature Uniformity @ -70°C – 5.0°C (9 probes, 3″ from wall)

Recovery from Door Opening to Steady-State* – 26 min @ -80°C

Warm-Up Time from -80°C to -60°C – 3.8 hrs.

Steady-State Energy Consumption* – 5.8kWh/day at -75°C

Steady-State Heat Dissipation – 754 BTU/h @ -80°C

10″ Touchscreen

Onboard Predictive Analytics – Yes

BACnet I/P – Yes

External Communications – Dry Contacts, 420mA

Engine – M6D

Storage Capacity – 795L

Temperature Uniformity @ -80°C – 6.0°C (20 probes, 1″ from wall)

Temperature Uniformity @ -70°C – 4.0°C (20 probes, 1″ from wall)

Recovery from Door Opening to Steady-State* – 26 min @ -80°C

Warm-Up Time from -80°C to -60°C – 3.8 hrs.

Steady-State Energy Consumption* – 6.6kWh/day at -75°C

Steady-State Heat Dissipation – 836 BTU/h @ -80°C

10″ Touchscreen

Onboard Predictive Analytics – Yes

BACnet I/P – Yes

External Communications – Dry Contacts, 420mA

Engine – M6D with Quick Change Technology

* ENERGY STAR Final Test Method: 15s door opening, 1 inner door

English (United States)